ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000
Erbe USA Inc
The hazard
A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture. FDA Class II recall.
The remedy
Follow FDA guidance