Food and Drug Administration · Recall Z-2938-2026 · Jun 26, 2026

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

Erbe USA Inc

The hazard

A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture. FDA Class II recall.

The remedy

Follow FDA guidance

About this recall

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

Affected identifiers

Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.

UPC 04050147013797