Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO4010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
TISSIUM SA
The hazard
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery. FDA Class II recall.
The remedy
Follow FDA guidance
About this recall
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO4010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
Affected identifiers
Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.
Model Lot Code: Model No NCOCO4010Model UDI-DI 03760279380148Model Lot Numbers 251000041Model 250700048Model 260300046Model 260400050Model and 260300047UPC 03760279380148Would you have heard about this one?
RapidRecall watches everything you own against every new federal recall and emails you the moment something you bought turns up. Free for your most important items.
Start monitoring freeSourced from the Food and Drug Administration’s public recall data. Always verify serial and date ranges against the official notice before acting.