All recalls
FDA
Food and Drug Administration · Recall Z-2904-2026 · Jun 15, 2026

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

TISSIUM SA

The hazard

Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery. FDA Class II recall.

The remedy

Follow FDA guidance

About this recall

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

Affected identifiers

Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.

Model Lot Code: Model No NCOCO2075Model UDI-DI ¿03760279380100Model Lot Number 250700016Model 251000048Model 251000049Model and 260300043UPC 03760279380100

Would you have heard about this one?

RapidRecall watches everything you own against every new federal recall and emails you the moment something you bought turns up. Free for your most important items.

Start monitoring free