Food and Drug Administration · Recall Z-2735-2026 · Jun 25, 2026

Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.

ResMed Ltd.

The hazard

Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy. FDA Class I recall.

The remedy

Follow FDA guidance

About this recall

Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.

Affected identifiers

Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.

UPC 00619498003259UPC 00619498270033