Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device
Baxter Healthcare Corporation
The hazard
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup. FDA Class I recall.
The remedy
Follow FDA guidance
About this recall
Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device
Affected identifiers
Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.
Model UDI: 00887761981492UPC 00887761981492Would you have heard about this one?
RapidRecall watches everything you own against every new federal recall and emails you the moment something you bought turns up. Free for your most important items.
Start monitoring freeSourced from the Food and Drug Administration’s public recall data. Always verify serial and date ranges against the official notice before acting.