All recalls
FDA
Food and Drug Administration · Recall Z-2548-2026 · May 21, 2026

Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device

Baxter Healthcare Corporation

The hazard

There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup. FDA Class I recall.

The remedy

Follow FDA guidance

About this recall

Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device

Affected identifiers

Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.

Model UDI: 00887761985018UPC 00887761985018

Would you have heard about this one?

RapidRecall watches everything you own against every new federal recall and emails you the moment something you bought turns up. Free for your most important items.

Start monitoring free

Sourced from the Food and Drug Administration’s public recall data. Always verify serial and date ranges against the official notice before acting.