MEDLINE Medical Procedure Kits labeled as: PELVIC LAP PACK, Medline Kit Number/ SKU DYNJ53820B
Medline Industries, LP
The hazard
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification. FDA Class II recall.
The remedy
Follow FDA guidance
About this recall
MEDLINE Medical Procedure Kits labeled as: PELVIC LAP PACK, Medline Kit Number/ SKU DYNJ53820B
Affected identifiers
Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.
Model UDI/DI 10195327548261 (each)Model 40195327548262 (case). Lot Numbers: 25LBI062Model 25HBR967Model 24JBK618Model 24BBQ860.UPC 10195327548261UPC 40195327548262Would you have heard about this one?