Food and Drug Administration · Recall Z-2486-2026 · May 4, 2026

MEDLINE Medical Procedure Kits labeled as: 1) BASIC HEART PACK, Medline Kit/SKU DYNJ67331B; 2) CARDIAC SURGERY, Medline Kit/SKU DYNJ911589A; 3) CAROTID ENDARTECOMY, Medline Kit/SKU DYNJ36211D; 4) CHLA ECMO / CARDIAC PACK, Medline Kit/SKU DYNJ87973; 5) CHLA ECMO / CARDIAC PACK, Medline Kit/SKU DYNJ87

Medline Industries, LP

The hazard

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification. FDA Class II recall.

The remedy

Follow FDA guidance

About this recall

MEDLINE Medical Procedure Kits labeled as: 1) BASIC HEART PACK, Medline Kit/SKU DYNJ67331B; 2) CARDIAC SURGERY, Medline Kit/SKU DYNJ911589A; 3) CAROTID ENDARTECOMY, Medline Kit/SKU DYNJ36211D; 4) CHLA ECMO / CARDIAC PACK, Medline Kit/SKU DYNJ87973; 5) CHLA ECMO / CARDIAC PACK, Medline Kit/SKU DYNJ87973A; 6) CV PART 1 AND 2, Medline Kit/SKU DYNJ908879; 7) CV PART 1 AND 2, Medline Kit/SKU DYNJ908879C; 8) FEM POP CDS, Medline Kit/SKU CDS983670K; 9) LAP BASIN TRAY, Medline Kit/SKU DYNJ50908B; 10) OPEN HEART LTX CATHETERS KIT, Medline Kit/SKU DYNJ0049032A; 11) OPEN HEART PACK, Medline Kit/SKU DYNJ80549C; 12) OPEN HEART PACK, Medline Kit/SKU DYNJ80549D; 13) RR-ADULT ECMO PACK, Medline Kit/SKU DYNJ69742B; 14) TAVR, Medline Kit/SKU DYNJ906509C.

Affected identifiers

Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.

Model 40195327013838 (case)Model Lot Number 23JBM539Model Lot Number 23IBF088Model Lot Number 23FBM203Model Lot Number 22JBK233Model Lot Number 22JBB115Model Lot Number 22IBO304Model Lot Number 22HBR331Model Lot Number 22FBM460Model Lot Number 22DBS093Model Lot Number 22CBJ072Model Lot Number 22BBX288Model Lot Number 21LBU240UPC 10195327013837UPC 40195327013838