All recalls
FDA
Food and Drug Administration · Recall Z-2463-2026 · May 29, 2026

Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Hamilton Medical AG

The hazard

Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement FDA Class I recall.

The remedy

Follow FDA guidance

About this recall

Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Affected identifiers

Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.

Model REF/UDI-DI: 260127/07630002802956Model 17630002802953Model 260128/07630002802963Model 17630002802960Model 260167/07630002802970Model 17630002802977Model 260168/07630002802987UPC 07630002802956UPC 17630002802953UPC 07630002802963UPC 17630002802960UPC 07630002802970UPC 17630002802977UPC 07630002802987UPC 17630002802984

Would you have heard about this one?

RapidRecall watches everything you own against every new federal recall and emails you the moment something you bought turns up. Free for your most important items.

Start monitoring free

Sourced from the Food and Drug Administration’s public recall data. Always verify serial and date ranges against the official notice before acting.