Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Hamilton Medical AG
The hazard
Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement FDA Class I recall.
The remedy
Follow FDA guidance
About this recall
Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Affected identifiers
Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.
Model REF/UDI-DI: 260127/07630002802956Model 17630002802953Model 260128/07630002802963Model 17630002802960Model 260167/07630002802970Model 17630002802977Model 260168/07630002802987UPC 07630002802956UPC 17630002802953UPC 07630002802963UPC 17630002802960UPC 07630002802970UPC 17630002802977UPC 07630002802987UPC 17630002802984Would you have heard about this one?