Food and Drug Administration · Recall Z-2358-2026 · May 8, 2026

Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.

Datex-Ohmeda Inc

The hazard

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient. FDA Class I recall.

The remedy

Follow FDA guidance

About this recall

Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.

Affected identifiers

Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.

UPC 00813841020010