All recalls
FDA
Food and Drug Administration · Recall Z-0356-2027 · Oct 2, 2026

Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.

CORIN MEDICAL, LTD.

The hazard

Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments. FDA Class II recall.

The remedy

Follow FDA guidance

About this recall

Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.

Affected identifiers

Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.

Model Part Number: 1232040. UDI-DI/GTINModel 05056139239656. Lot Code: 097921.UPC 05056139239656