Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.
CORIN MEDICAL, LTD.
The hazard
Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments. FDA Class II recall.
The remedy
Follow FDA guidance
About this recall
Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.
Affected identifiers
Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.
Model Part Number: 1232040. UDI-DI/GTINModel 05056139239656. Lot Code: 097921.UPC 05056139239656Would you have heard about this one?
RapidRecall watches everything you own against every new federal recall and emails you the moment something you bought turns up. Free for your most important items.
Start monitoring freeSourced from the Food and Drug Administration’s public recall data. Always verify serial and date ranges against the official notice before acting.