Food and Drug Administration · Recall Z-0353-2027 · Sep 29, 2026

Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633

Baxter Healthcare Corporation

The hazard

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury. FDA Class II recall.

The remedy

Follow FDA guidance

About this recall

Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633

Affected identifiers

Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.

Model Serial numbers: 108920104Model 110192607Model 110192885Model 110208936Model 110212885Model 110220066Model 110227520Model 110248337Model 110250614Model 109423464Model 110192608Model 110192886Model 110208941Model 110212886Model 110220068Model 110227523Model 110248343Model 110250621Model 109517138Model 110192609Model 110206656Model 110210717Model 110220055Model 110220069Model 110227524UPC 00887761967519

Would you have heard about this one?

RapidRecall watches everything you own against every new federal recall and emails you the moment something you bought turns up. Free for your most important items.

Start monitoring free

Sourced from the Food and Drug Administration’s public recall data. Always verify serial and date ranges against the official notice before acting.