Operating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788
Baxter Healthcare Corporation
The hazard
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury. FDA Class II recall.
The remedy
Follow FDA guidance
About this recall
Operating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788
Affected identifiers
Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.
Model Serial numbers: 108517025Model 110115554Model 110144247Model 110188916Model 109558596Model 110139748Model 110166201Model 110188918Model 109737582Model 110139769Model 110177678Model 110192686Model 110115551Model 110144245Model 110185616Model 109889060Model UDI-DI number: 00887761967526UPC 00887761967526Would you have heard about this one?
RapidRecall watches everything you own against every new federal recall and emails you the moment something you bought turns up. Free for your most important items.
Start monitoring freeSourced from the Food and Drug Administration’s public recall data. Always verify serial and date ranges against the official notice before acting.