Food and Drug Administration · Recall Z-0351-2027 · Sep 29, 2026

Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695

Baxter Healthcare Corporation

The hazard

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury. FDA Class II recall.

The remedy

Follow FDA guidance

About this recall

Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695

Affected identifiers

Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.

Model Serial numbers: 108094281Model 108468075Model 108484417Model 108506318Model 108516636Model 108536785Model 108544115Model 108604154Model 108463223Model 108484404Model 108489634Model 108511411Model 108516637Model 108544106Model 108544116Model 108618964Model 108463224Model 108484405Model 108489648Model 108511412Model 108521436Model 108544107Model 108604137Model 108618965Model 108468062UPC 00887761974425

Would you have heard about this one?

RapidRecall watches everything you own against every new federal recall and emails you the moment something you bought turns up. Free for your most important items.

Start monitoring free

Sourced from the Food and Drug Administration’s public recall data. Always verify serial and date ranges against the official notice before acting.