Food and Drug Administration · Recall Z-0350-2027 · Sep 29, 2026

TruSystem 7500 surgical table, Model/Catalog Number: 4091000

Baxter Healthcare Corporation

The hazard

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury. FDA Class II recall.

The remedy

Follow FDA guidance

About this recall

TruSystem 7500 surgical table, Model/Catalog Number: 4091000

Affected identifiers

Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.

Model Serial numbers: 108491706Model 108726496Model 108758657Model 108858029Model 108945825Model 109083212Model 109343910Model 109684319Model 110016886Model 108557062Model 108736192Model 108773032Model 108882327Model 108976094Model 109174333Model 109370104Model 109690845Model 110016887Model 108574826Model 108736193Model 108773033Model 108882328Model 108976095Model 109174334Model 109396256UPC 00887761968639

Would you have heard about this one?

RapidRecall watches everything you own against every new federal recall and emails you the moment something you bought turns up. Free for your most important items.

Start monitoring free

Sourced from the Food and Drug Administration’s public recall data. Always verify serial and date ranges against the official notice before acting.