All recalls
FDA
Food and Drug Administration · Recall Z-0348-2027 · Oct 1, 2026

Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner

restor3d Inc.

The hazard

(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery. FDA Class II recall.

The remedy

Follow FDA guidance

About this recall

Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner

Affected identifiers

Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.

Model Product Code: 3500-140U3 UDI/GTIN:00840097513461Model Product Code: 3500-140C1 UDI/GTIN:00840097513430Model Product Code: 3500-140U1 UDI/GTIN:00840097513409UPC 00840097513492UPC 00840097513461UPC 00840097513430UPC 00840097513409

Would you have heard about this one?

RapidRecall watches everything you own against every new federal recall and emails you the moment something you bought turns up. Free for your most important items.

Start monitoring free

Sourced from the Food and Drug Administration’s public recall data. Always verify serial and date ranges against the official notice before acting.