Embletta MPR PG Multi Parameter Recorder Sleep Monitor, Model/Catalog Number: 2003011

NATUS MEDICAL INCORPORATED

The hazard

The device may be unable to correctly format its internal memory. When the internal memory cannot be correctly formatted, previously recorded studies may remain on the device and may be updated with incorrect patient information from a subsequent study. The issue may also prevent studies from initiating recording successfully or may cause recordings to end prematurely, resulting in incomplete or unavailable diagnostic data and the potential need to repeat a patient study. FDA Class II recall.

The remedy

Follow FDA guidance

About this recall

Embletta MPR PG Multi Parameter Recorder Sleep Monitor, Model/Catalog Number: 2003011

Affected identifiers

Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.

Model UDI-DI: 00382830023764UPC 00382830023764