Carevo, Height-adjustable shower trolley, Model Number BAC1111-01

ARJOHUNTLEIGH POLSKA Sp. z.o.o.

The hazard

The side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient handling. If the side support is not fully locked and an additional force is applied by the caregiver on the handle (e.g., during turning or repositioning), the panel may open unintentionally, which can lead to a resident/patient fall and serious injury, such as head injury, fractures, or other trauma resulting from a patient fall. FDA Class II recall.

The remedy

Follow FDA guidance

About this recall

Carevo, Height-adjustable shower trolley, Model Number BAC1111-01