All recalls
FDA
Food and Drug Administration · Recall Z-0298-2027 · Oct 6, 2026

Philips Azurion 7 M12. Model Numbers: 722078, 722223.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

The hazard

Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images. FDA Class II recall.

The remedy

Follow FDA guidance

About this recall

Philips Azurion 7 M12. Model Numbers: 722078, 722223.

Affected identifiers

Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.

Model (01)00884838085251(21)585Model (01)00884838085251(21)293Model (01)00884838085251(21)613Model (01)00884838085251(21)546Model (01)00884838085251(21)479Model (01)00884838085251(21)278Model (01)00884838085251(21)456Model (01)00884838085251(21)571Model (01)00884838085251(21)921Model (01)00884838085251(21)165Model (01)00884838085251(21)656Model (01)00884838085251(21)657Model (01)00884838085251(21)576Model (01)00884838085251(21)121Model (01)00884838085251(21)383Model (01)00884838085251(21)239Model (01)00884838085251(21)645Model (01)00884838085251(21)379Model (01)00884838085251(21)399Model (01)00884838085251(21)803Model (01)00884838085251(21)296Model (01)00884838085251(21)559Model (01)00884838085251(21)538Model (01)00884838085251(21)166Model (01)00884838085251(21)257UPC 00884838085251

Would you have heard about this one?

RapidRecall watches everything you own against every new federal recall and emails you the moment something you bought turns up. Free for your most important items.

Start monitoring free

Sourced from the Food and Drug Administration’s public recall data. Always verify serial and date ranges against the official notice before acting.