Philips Azurion 7 B12. Model Numbers: 722067, 722225.
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
The hazard
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images. FDA Class II recall.
The remedy
Follow FDA guidance
About this recall
Philips Azurion 7 B12. Model Numbers: 722067, 722225.
Affected identifiers
Check these against the label on your unit. An identifier match is how a recall is confirmed, not just suspected.
Model (01)00884838085350(21)173Model (01)00884838085350(21)172Model (01)00884838085350(21)109Model (01)00884838085350(21)250Model (01)00884838085350(21)114Model (01)00884838085350(21)69Model (01)00884838085350(21)214Model (01)00884838085350(21)221Model (01)00884838085350(21)81Model (01)00884838085350(21)278Model (01)00884838085350(21)131Model (01)00884838085350(21)141Model (01)00884838085350(21)26Model (01)00884838085350(21)133Model (01)00884838085350(21)128Model (01)00884838085350(21)151Model (01)00884838085350(21)68Model (01)00884838085350(21)233Model (01)00884838085350(21)255Model (01)00884838085350(21)33Model (01)00884838085350(21)256Model (01)00884838085350(21)270Model (01)00884838085350(21)238Model (01)00884838085350(21)187Model (01)00884838085350(21)74UPC 00884838085350Would you have heard about this one?
RapidRecall watches everything you own against every new federal recall and emails you the moment something you bought turns up. Free for your most important items.
Start monitoring freeSourced from the Food and Drug Administration’s public recall data. Always verify serial and date ranges against the official notice before acting.